A NEW SIMPLE METHOD FOR ACCURATE MEASUREMENT OF URINARY I<sup>131</sup>AFTER TRACER AND THERAPEUTIC DOSES*

A. STONE FREEDBERG, Alvin L. Ureles, MARVIN VAN DILLA, MARY J. McMANUS

The Journal of Clinical Endocrinology & Metabolism · 1950 · 13 citations · 1 references

Concepts

Abstract

WE HAVE recently (1) described a method for quantitative measurement of the amount of I131 in the thyroid gland by external counting. It was apparent that the method was not only applicable to the external counting of I131 in the thyroid but likewise afforded a simple, direct measurement of I131 content in urine, as shown by the experimental data presented here. This new method of measurement is as accurate as the standard gamma counting technique previously reported from this laboratory, is simple, and requires minimum handling and exposure of the urine sample (2) Four platinum cathode Geiger-Mueller tubes were connected in a parallel electrical circuit and arranged in a horizontal plane (Fig. 1). The distance of each tube from the center was 60.0 cm. We have previously shown (3) that within a “circle” of this diameter (120.0 cm.) is a smaller concentric sphere, approximately 40 cm. in diameter, in which a source of I131 may be moved with a change in counting rate of not more than 5 percent, as compared to net counts per minute obtained with the I131 source in the center of the circle. We have termed this central space the “free area” and have shown that its diameter is proportional to the diameter of the Geiger-Mueller tube circle (Fig 2).

References