Radiation Research · 1997 · 87 citations · 13 references
EngineeringBiomedical EngineeringOncologyBiomedical OpticsBiophotonicsSensorsFiber-optic Medical DevicesOptical BiosensorsOptical DiagnosticsOptical ImagingOptical SensorsTumor ImagingHypoxic Cell FractionOxygen SensingOxygen SensorsOxygenationSensingTissue Oxygenation MeasurementTissue Oxygenation MonitoringTumor Oxygenation
A novel oxygen sensor which does not rely on electrochemical reduction has been used to measure the oxygenation of the murine sarcoma F in a comparative study with an existing polarographic electrode that is available commercially. The prototype luminescence sensor yielded an oxygen distribution comparable with readings made using a pO2 histograph. The percentage of regions detected that had a pO2 less than 5 mm Hg was 79 and 75 using the Eppendorf pO2 histograph and the luminescence fiber optic sensor, respectively. These values were compatible with a measured radiobiologically hypoxic fraction of 67% in this tumor. The polarographic method detected more regions with a pO2 of 2.5 mm Hg or less (69%) compared with the optical sensor (50%) (P < 0.05). This could reflect differences in the oxygen use of the sensing devices. This initial assessment indicates the potential of a fiber-optic-based oxygen-monitoring system. Such a system should have several advantages including monitoring temporal oxygen changes in a given microregion and use with NMR procedures.
13
Tumor oxygenation predicts for the likelihood of distant metastases in human soft tissue sarcoma.
David M. Brizel, Sean P. Scully, John M. Harrelson et al. · PubMed · 1996 · 1.2K citations