Artificial Organs · 2003 · 60 citations · 11 references
EngineeringBiomedical EngineeringCardiologyClinical CardiologyComputational Fluid DynamicsFluid DynamicsBiofluid MechanicsCardiovascular BiomechanicsCardiovascular System ModelingFlow ModelingHemodynamicsHemodynamics ModelingAxial-flow PumpsBlood Flow DynamicsCardiovascular Fluid DynamicsExposure Time DependencePressure - Flow CharacteristicsPump Flow DynamicsReynolds Stress ModelingShear Stress
A low rate of hemolysis is an important factor for the development of a rotary blood pump. It is, however, difficult to identify the areas where hemolysis occurs. Computational fluid dynamics (CFD) analysis enables the engineer to predict hemolysis on a computer. In this study, fluid dynamics throughout intracardiac axial flow pumps with different designs were analyzed three-dimensionally using CFD software. The computed pressure-flow characteristics of the pump were in good agreement with the measurements. The Reynolds shear stress was computed along particle trace lines. Hemolysis was estimated on the basis of shear stress (tau) and its exposure time (Deltat): dHb/Hb = 3.62 x 10(-7)(tau)(i)(2.416) x Delta(t)(i)(0.785). Particle damage increased with time along the particle trace lines. Hemolysis of each of the pumps was measured in vitro. The computed hemolysis values were in good agreement with the experimental results. CFD is a useful tool for developing a rotary blood pump.
11
A Mathematical Model for Shear‐Induced Hemolysis
Krishna K. Yeleswarapu, James F. Antaki, Marina V. Kameneva et al. · Artificial Organs · 1995 · 151 citations