The performance of MatriXX has been systematically evaluated in the peripheral dose regions. Major sources of error associated with MatriXX are identified and a correction method is suggested. This method has been successfully tested using both experimental and clinical plans. In all cases, good agreements between MatriXX and an ion chamber are achieved after corrections. The authors conclude that with proper corrections, MatriXX can be reliably used for peripheral dose measurements within the ranges studied.