High-Precision Atmospheric <sup>14</sup>CO<sub>2</sub> Measurement at the Rafter Radiocarbon Laboratory

Jocelyn Turnbull, Albert Zondervan, Johannes W. Kaiser, Margaret Norris, Jenny Dahl, Scott J. Lehman

Radiocarbon · 2015 · 35 citations · 20 references

Concepts

Abstract

This article describes a new capability for high-precision 14 C measurement of CO 2 from air at the Rafter Radiocarbon Laboratory, GNS Science, New Zealand. We evaluate the short-term within-wheel repeatability and long-term between-wheel repeatability from measurements of multiple aliquots of control materials sourced from whole air. Samples are typically measured to 650,000 14 C counts, providing a nominal accelerator mass spectrometry (AMS) statistical uncertainty of 1.3‰. No additional uncertainty is required to explain the within-wheel variability. An additional uncertainty factor is needed to explain the long-term repeatability spanning multiple measurement wheels, bringing the overall repeatability to 1.8‰, comparable to other laboratories measuring air materials to high precision. This additional uncertainty factor appears to be due to variability in the measured 14 C content of OxI primary standard targets, likely from the combustion process. We observe an offset of 1.4‰ in our samples relative to those measured by the University of Colorado INSTAAR, comparable to interlaboratory offsets observed in recent intercomparison exercises.

References

20