Physics in Medicine and Biology · 1996 · 31 citations · 4 references
MedicineRadiologyRadiation DosimetryRadiation MeasurementRadiation PhysicsAccelerator PhysicsExternal Beam DosimetryPhantom DosimetryPhoton DosimetryRadiation Dose MeasurementSmall Field DosimetryAcceleratorsClinical Photon Beam DosimetryLinear AcceleratorsParticle Beam DosimetryPhoton Beam QualityScatter Radiation CharacterizationScatter Radiation Dosimetry
Phantom scatter factors, for square fields of various sizes, have been determined at a fixed reference depth of 10 cm, separately in different institutes, for a large number of linear accelerators under the auspices of the Netherlands Commission on Radiation Dosimetry. The method used for these measurements has been described in a previous paper. The present article describes the conversion of the measured values into a comprehensive and consistent data set, that gives the phantom scatter factor as a function of field size (from 4 cm up to 40 cm) and quality index (from 0.600 up to 0.800).
4
Intercomparison of normalized head-scatter factor measurement techniques
Douglas M. D. Frye, Bhudatt R. Paliwal, Bruce Thomadsen et al. · Medical Physics · 1995 · 31 citations · Full text