Accuracy of Transcutaneous Carbon Dioxide Tension Measurements during Cardiopulmonary Exercise Testing

Gerben Stege, Frank J.J. van den Elshout, Yvonne F. Heijdra, Marjo J.T. van de Ven, P.N.R. Dekhuijzen, Petra J.E. Vos

Respiration · 2008 · 35 citations · 12 references

Concepts

Abstract

<i>Background:</i> Measurements of transcutaneous carbon dioxide tension (Ptc<i>CO</i><sub>2</sub>) with current devices are proven to provide clinically acceptable agreement with measurements of partial arterial carbon dioxide tension (Pa<i>CO</i><sub>2</sub>) in several settings but not during cardiopulmonary exercise testing (CPET). <i>Objectives:</i> The primary objective of this study was to investigate the agreement between Pa<i>CO</i><sub>2</sub> and Ptc<i>CO</i><sub>2</sub> measurements (using a Tosca 500 with a Tosca sensor 92) during CPET. A secondary objective was to investigate the agreement between arterial and transcutaneous oxygen saturation (Sa<i>O</i><sub>2</sub>, Sp<i>O</i><sub>2</sub>) as measured with this sensor during CPET. <i>Methods:</i> In patients with various pulmonary diseases, Ptc<i>CO</i><sub>2</sub> and Sp<i>O</i><sub>2</sub> were continuously measured and compared with arterial blood gas samples during CPET. A maximum bias of 0.5 kPa and 95% limits of agreement (LOA) of 1 kPa between carbon dioxide pressure (P<i>CO</i><sub>2</sub>) measurements were determined as clinically acceptable. <i>Results:</i> In total 101 ‘paired’ arterial and transcutaneous measurements were obtained from 21 patients. Bias between Pa<i>CO</i><sub>2</sub> and Ptc<i>CO</i><sub>2</sub> was –0.03 kPa with LOA from –0.78 to 0.71 kPa. Bias between Sa<i>O</i><sub>2</sub> and Sp<i>O</i><sub>2</sub> was –1.0% with LOA from –2.83 to 0.83%. <i>Conclusions:</i> Transcutaneous estimations of P<i>CO</i><sub>2</sub> and Sp<i>O</i><sub>2</sub> are accurate and can be used in CPET, circumvening the need for arterial cannulation.

References

12