Occupational exposure to static and time‐varying gradient magnetic fields in MR units

Jill K. Bradley, Matilda Nyekiova, David L. Price, Lindsay D'jon Lopez, Therese Crawley

Journal of Magnetic Resonance Imaging · 2007 · 42 citations · 3 references

DOIFull text

Open access

Concepts

Abstract

PURPOSE: To assess exposure of MR staff to static and time-varying gradient magnetic fields by direct measurement and questionnaire and to compare these to the action values and exposure limits stated in published guidance. MATERIALS AND METHODS: Instantaneous and time-averaged exposure to static magnetic fields were measured on clinical and research MR scanners. A questionnaire was sent to 100 clinical MR centers to estimate the average static field exposure. Instantaneous measurements of the time-varying magnetic field were conducted on a 1.5 T conventional and a 0.6 T open-bore MR system. RESULTS: The 24-hour time-averaged static field exposure was measured to be 1.8 +/- 1 mT and the questionnaire produced a 24-hour time-average of 2.3 mT. The maximum static field strength recorded during patient set up was 554 mT. Time-varying gradient field exposure was found to exceed the action values up to 30 cm from the bore entrance or magnet rim. CONCLUSION: The time-varying gradient field exposure limit is the only limit to be exceeded. The time-averaged static field exposure to staff was found to be 100 times lower than the recommended guidelines.