IEEE Transactions on Instrumentation and Measurement · 1964 · 52 citations · 6 references
EngineeringElectrical EngineeringInstrumentationMicrowave EngineeringThermal AnalysisCalibrationEquipment CalibrationMicrowave MeasurementsMicrowave Noise AnalysisNoiseNoise MeasurementExcess NoiseMicrowave Noise CharacterizationNoise CalibrationNoise Source CharacterizationNoise SourcesSignal-to-noise RatioSystem ComparisonTest Frequencies
This report describes a system now in operation for the purpose of calibrating microwave noise sources at three selected frequencies in the range 8.2 to 12.4 Gc. Included are a discussion of the reference standard noise source, an error analysis of the standard source and the comparison system, and evidence of system performance. Results of measurements indicate that the excess noise ratio of a specific commercial argon filled noise source is 15.6 db at 9.8 Gc. A calibration service is now available for tube-in-mount noise sources at the three selected frequencies 9.0 Gc, 9.8 Gc, and 11.2 Gc.
6
Absolute Measurement of Temperatures of Microwave Noise Sources
A. J. Estin, C. L. Trembath, J. S. Wells et al. · IRE Transactions on Instrumentation · 1960 · 19 citations