Publication | Open Access
Rapid Detection of Myocardial Infarction With a Sensitive Troponin Test
18
Citations
17
References
2011
Year
Heart FailureCardiovascular DiseaseSensitive Troponin AssayAtherosclerosisMyocardial InfarctionCoronary Artery DiseaseDiagnosisLaboratory MedicineBiostatisticsAcute Myocardial InfarctionPublic HealthMedicineCardiologyEmergency DepartmentEmergency MedicineRadiologyCardiovascular Imaging
Rapid identification and treatment of patients with a myocardial infarction (MI) is mandatory. We studied the diagnostic capacities of a sensitive troponin assay for detection of MI in emergency department patients within 2 hours after arrival. The study included 157 patients suspected of having non-ST-elevation acute coronary syndrome. Blood was drawn on arrival (T0) and 2 (T2), 6, and 12 hours later. At T2, a troponin concentration above the MI cutoff is 87% sensitive and 100% specific for MI detection (positive predictive value [PPV], 100%; negative predictive value [NPV], 96%). If a difference of more than 30% between the troponin measurements at T0 and T2 in the absence of an absolute troponin increase above the 99th percentile of a reference population is also considered indicative of MI, the sensitivity increases to 100% and specificity decreases to 87% (PPV, 70%; NPV, 100%). Sensitivity and specificity of creatine kinase-MB and myoglobin are lower than those of troponin. By using a sensitive troponin assay and simple algorithms, the diagnosis of MI can be determined within 2 hours after arrival at the emergency department. Measurement of myoglobin and creatine kinase-MB has no added value.
| Year | Citations | |
|---|---|---|