Undersensing of P Waves in the Presence of an Adequate P Wave Due to Automatic Postventricular Atrial Refractory Period Extension

John H. Wilson, STEPHEN E. LATTNER

Pacing and Clinical Electrophysiology · 1989 · 17 citations · 3 references

Concepts

Abstract

WILSON, J.H., et al .: Undersensing of P Waves in the Presence of an Adequate P Wave Due to Automatic Postventricular Atrial Refractory Period Extension Two cases in which a DDD pacemaker failed to track P waves due to repetitive activation of the automatic postventricular atrial refractory period (PVARP) extension are reported. In one case, the cycle was initiated by a premature ventricular contraction (PVC); in the other, by touching the unipolar pacemaker with a needle while suturing. We demonstrated that chest wall stimulation can be used to induce cycles of repetitive automatic PVARP extension and failure to track P waves. We demonstrated in vitro that touching a unipolar pacemaker with a needle can generate a signal suffiicient to mimic a QRS complex, and initiate a cycle of repetitive P wave undersensing.

References

3