An ac System for Measuring Thermopower

Ross H. Freeman, J. Bass

Review of Scientific Instruments · 1970 · 33 citations · 2 references

Concepts

Abstract

Using available, or easily borrowed, equipment we have constructed an ac system for measuring thermopower. With this system we have measured the thermopower of a gold-aluminum thermocouple from 30 to 150°C with accuracy of a few percent, using a temperature difference of 0.03 K. With care, and the proper equipment, measurements accurate to about 1% appear possible, using a temperature difference smaller than 0.01 K.