IEEE Transactions on Nuclear Science · 2007 · 18 citations · 9 references
EngineeringElectrical EngineeringMicroelectronicsRadiation DetectionRadiation MeasurementDetector SystemsDetector TechnologyPhotodetectorsRadiation DetectorsSemiconductor DetectorsDetector AgingDetector DesignDetector Energy ResolutionDetectorsPhotodetector DetectivityPolarizationRadiation Energy ResolutionSemiconductor Detector Resolution
The energy resolution of CdTe detectors typically deteriorates in several tens of minutes due to the polarization effect at 30degC. In this paper, the energy resolutions were evaluated at 30degC and 35degC, and the periodic bias reset technique was investigated as a way to stabilize the detector performance. Two different bias reset circuits were used, one was an exponential voltage change type and the other was a linear voltage change type. The 500 V bias with 0.5 s to 60 s reset durations and 1 min to 20 min reset intervals was applied to 1 mm thick CdTe detectors. It was found that the 0.5 s reset duration and 5 min reset interval were able to stabilize the detector performance at 35degC.
9
Polarization in Cadmium Telluride Nuclear Radiation Detectors
P. Siffert, JA Berger, C. Scharager et al. · IEEE Transactions on Nuclear Science · 1976 · 147 citations
Natural Sciences, Materials Science, Nuclear Engineering +14
High-resolution Schottky CdTe diode for hard X-ray and gamma-ray astronomy
Tadayuki Takahashi, Biswajit Paul, Kazuyuki Hirose et al. · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · 1999 · 145 citations · Full text
Natural Sciences, Electrical Engineering, Materials Science +12
High-resolution Schottky CdTe diode detector
Tadayuki Takahashi, Takefumi Mitani, Y. Kobayashi et al. · IEEE Transactions on Nuclear Science · 2002 · 104 citations