Publication | Closed Access
Evaluation of a lung tumor autocontouring algorithm for intrafractional tumor tracking using low-field MRI: A phantom study
37
Citations
36
References
2012
Year
Dice's coefficients of > 0.96 and > 0.93 are achieved between autocontoured and real tumor shapes, and the position of a tumor can be tracked with RMSE values of < 0.55 and < 0.92 mm in 0.5 and 0.2 T equivalent images, respectively. These results demonstrate the feasibility of lung tumor autocontouring in low field MR images, and, by extension, intrafractional lung tumor tracking with our laboratory's linac-MR system.