Design and implementation of a digital filter for beat-by-beat impedance cardiography

Y. Yamamoto, K. Mokushi, Shigeaki Tamura, Yoshiteru Mutoh, M. Miyashita, Hirotsugu Hamamoto

IEEE Transactions on Biomedical Engineering · 1988 · 32 citations · 20 references

Concepts

Abstract

A digital filter for thoracic impedance cardiography was developed and implemented on a 16-bit personal computer after examining the effect of respiratory movement on the first derivative of the thoracic impedance signal. Four male subjects exercised with a cycle ergometer at 100 and 150 W successively, after resting for 5 min. Thoracic impedance and its first derivative (dZ/dt) were recorded by a standard four-electrode cardiograph. The peak-power spectral densities of the pneumogenic (P/sub p,c/) and the cardiogenic (P/sub p,c/) components of dZ/dt were separated with a simultaneous recording of the ECG and thoracic circumferences. P/sub p/P/sub c/ increased with each increment of work rate: 0.12 (0.05-0.19) at rest, 0.67 (0.49-0.97) at 100 W, and 0.97 (0.58-1.52) at 150 W.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">&gt;</ETX>

References

20