Clinical Physics and Physiological Measurement · 1980 · 13 citations · 2 references
MedicineDermatologyBiomedical OpticsRadiation DosimetryRadiation MeasurementDose CalibrationPhoton DosimetryPsoralen Plus Ultraviolet A TherapyRadiation Dose MeasurementUltraviolet InstrumentationAbsolute CalibrationAngular ResponseHigh-risk Cutaneous CarcinomaIrradianceLight DosimetryRadiation Dosimetry CalibrationUltraviolet Radiation MeasurementUva Dosimetry
The absolute calibration, linearity, angular response and wavelength response of three different commercially available UVA meters have been evaluated. The meters exhibited wide variations in accuracy in determining UVA irradiance, combined with significant differences in response at different angles and wavelengths. It is considered that these variations may lead to uncertainties in ultraviolet radiation dosimetry when assessing the relation between the risk of cutaneous carcinoma associated with oral psoralen photochemotherapy and UVA dose.
2