Scanning thermoacoustic tomography in biological tissue

Geng Ku, Lihong V. Wang

Medical Physics · 2000 · 161 citations · 32 references

Concepts

TL;DR

Microwave‑induced thermoacoustic tomography was explored to image biological tissue. Short microwave pulses induce thermoelastic expansion, generating acoustic waves detected by a focused ultrasonic transducer, and scanning perpendicular to the acoustic axis yields multi‑dimensional images while axial and lateral resolutions are characterized. Two‑dimensional tomographic images of biological tissue were obtained using 3‑GHz microwaves, and the simulated piezoelectric response matched the experimental signal closely.

Abstract

Microwave‐induced thermoacoustic tomography was explored to image biological tissue. Short microwave pulses irradiated tissue to generate acoustic waves by thermoelastic expansion. The microwave‐induced thermoacoustic waves were detected with a focused ultrasonic transducer. Each time‐domain signal from the ultrasonic transducer represented a one‐dimensional image along the acoustic axis of the ultrasonic transducer similar to an ultrasonic A‐scan. Scanning the system perpendicularly to the acoustic axis of the ultrasonic transducer would generate multi‐dimensional images. Two‐dimensional tomographic images of biological tissue were obtained with 3‐GHz microwaves. The axial and lateral resolutions were characterized. The time‐domain piezo‐electric signal from the ultrasonic transducer in response to the thermoacoustic signal was simulated theoretically, and the theoretical result agreed with the experimental result very well.

References