Concepedia

Publication | Closed Access

Evolution of the Ablation Region After Magnetic Resonance–Guided High-Intensity Focused Ultrasound Ablation in a Vx2 Tumor Model

32

Citations

26

References

2013

Year

Abstract

During the first 3 days after MR-HIFU ablation, the ablation region increases in size, after which it gradually decreases in size. The NPA on CE-T1-w imaging underestimates the extent of tissue necrosis on histology in the initial few days, but after 1 week, the NPA is reliable in delineating the necrotic tissue area. The 240-EM thermal dose limit underestimates the necrotic tissue area immediately after MR-HIFU ablation. Reliable treatment evaluation techniques are particularly important for noninvasive, image-guided tumor ablation. Our results indicate that CE-T1-w imaging is reliable for MR-HIFU treatment evaluation after 1 week.