Applied Physics Letters · 2006 · 49 citations · 11 references
EngineeringElectrical EngineeringPhysicsRadiation DetectionRadiation MeasurementRadiation PhysicsDetector TechnologyEnergy ResolutionGamma-ray DetectionGamma-ray DetectorsGamma-ray InstrumentationRadiation DetectorsDetector Energy ResolutionDetector NoiseDetector Noise CharacterizationDetectorsMicrocalorimeter DetectorsRadiation Energy ResolutionTransition Edge SensorsTransition-edge Sensor Detectors
The authors describe a microcalorimeter γ-ray detector with measured energy resolution of 42eV full width at half maximum for 103keV photons. This detector consists of a thermally isolated superconducting transition-edge thermometer and a superconducting bulk tin photon absorber. The absorber is attached with a technique compatible with producing arrays of high-resolution γ-ray detectors. The results of a detailed characterization of the detector, which includes measurements of the complex impedance, detector noise, and time-domain pulse response, suggest that a deeper understanding and optimization of the thermal transport between the absorber and thermometer could significantly improve the energy resolution of future detectors.
11
Thermal detectors as x-ray spectrometers
S. H. Moseley, John C. Mather, D. McCammon · Journal of Applied Physics · 1984 · 454 citations · Full text
Microcalorimeter and bolometer model