Comparison of Two Different Methods for Measurement of Phenylalanine in Dried Blood Spots

Adriana Rivero, J.A Allué, Ana Grijalba

Clinical Chemistry and Laboratory Medicine (CCLM) · 2000 · 17 citations · 8 references

Concepts

Abstract

Phenylketonuria is an inherited metabolic disorder caused by a defect in the hydroxylation of phenylalanine. Newborn screening is crucial for the diagnosis and treatment of this disease. A phenylalanine dehydrogenase-coupled enzymatic assay (Quantase) in microtiter plates for the screening of phenylketonuria was evaluated and compared with our routine method based on the modified fluorometric McCaman method. The test exhibited a linear calibration curve with a good slope as well as sufficient imprecision (< 10%), recovery (99.23+/-4.86%) and limit of detection (54.5 micromol/l). One hundred and ninety dried blood spots were analysed by this enzymatic method and compared with McCaman's. Although Quantase (Teknovas, Bilbao, Spain) showed a phenylalanine mean level in dried blood spot 18.2 micromol/l higher than that obtained with our routine method, the agreement between both techniques was considered acceptable.