American Journal of Hypertension · 1997 · 43 citations · 21 references
MedicinePhysiologyCircadian RhythmsSleep MedicineTime Series AnalysisBlood Pressure ControlCardiovascular MonitoringAgreement AnalysisBlood Pressure MonitoringBlood Pressure MonitorsBlood Pressure VariabilityBlood Pressure Variability AnalysisContinuous Blood Pressure Monitoring
We tested the reproducibility of ambulatory blood pressure monitoring (ABPM) by the use of agreement plots. Thirty-two normotensive volunteers underwent ABPM on four separate days (interval 28 days), on the same typical weekday. Sleeping time was restricted to the ABPM nighttime subperiod from 11:00 PM to 7:00 AM. Twenty-four-hour average values-both systolic and diastolic-daytime average values, and nighttime average values, as well as standard deviation (SD) values, were analyzed for differences (analysis of variance). Adaptation occurred from the first to the fourth ABPM, ie, average 24 h, daytime, and nighttime values were lower (-1 to -3 mm Hg) during the fourth recording than the first (P < .05 to P < .01). The agreement analysis showed a surprisingly high agreement among the four data sets (ie, differences from +/-2.54 to +/-5.92 mm Hg; +/-2 SD of the distribution). We concluded that reproducibility of ABPM seems excellent, but adaptation may occur, even in normotensive volunteers under research conditions. Caution must be paid before labeling a patient as hypertensive, because initial ABPM may yield higher values than later monitorings.
21
The Prognostic Value of Ambulatory Blood Pressures
Ambulatory blood pressure normality
Giuseppe Mancia, Roberto Sega, C. Bravi et al. · Journal of Hypertension · 1995 · 547 citations