Rapid single‐scan <i>T</i>‐mapping using exponential excitation pulses and image‐based correction for linear background gradients

Simon Baudrexel, Steffen Volz, Christine Preibisch, Johannes Klein, Helmuth Steinmetz, Rüdiger Hilker, Ralf Deichmann

Magnetic Resonance in Medicine · 2009 · 73 citations · 21 references

Concepts

TL;DR

The authors present a rapid quantitative T(2)* mapping technique that uses exponential excitation pulses and multi‑gradient‑echo imaging with static field inhomogeneity correction. Gradient maps derived from phase data correct modulus images, long‑TE echoes are omitted, and a monoexponential fit yields T(2)* values in 15 s per 256×256 slice, with additional in‑plane gradient correction improving accuracy. Phantom tests show accurate T(2)* values up to 200 µT/m gradients and errors below 15 % at 300 µT/m, while in‑vivo volunteer data at 3 T agree closely with literature values.

Abstract

A method for fast quantitative T(2)* mapping based on multiple gradient-echo (multi-GE) imaging with correction for static magnetic field inhomogeneities is described, using an exponential excitation pulse. Field gradient maps are obtained from the phase information and modulus data are subsequently corrected, allowing for simple monoexponential T(2)* fitting. Echoes with long echo times suffering from major signal losses due to field inhomogeneities are excluded from the analysis. The acquisition time for a matrix size of 256 x 256, 1 mm in-plane resolution, and 2 mm slice thickness amounts to 15 s per slice. An additional correction for in-plane field gradients further improves accuracy. Phantom experiments show that the method provides accurate T(2)* values for field gradients up to 200 microT/m; for gradients up to 300 microT/m errors do not exceed 15%. In vivo T(2)* values acquired on healthy volunteers at 3T are in excellent agreement with results from the literature.

References

21