British Journal of Ophthalmology · 1994 · 28 citations · 4 references
MedicineOphthalmologyVision ScienceClinical Vision ScienceContrast SensitivityVisual OpticsDynamic Visual AcuityOcular ImplantsPeripheral Contrast SensitivityVisual AcuityVisual ImpairmentAdaptation To Multifocal LensesContrast Sensitivity ImprovementLow-contrast Visual AcuityMesopic VisionMonofocal Intraocular LensesStatic Visual AcuityVisual Acuity GainsVisual Acuity Impairment
The functional results (contrast sensitivity and dynamic visual acuity) of 19 multifocal (3M design) and 14 all polymethylmethacrylate biconvex monofocal intraocular lenses (IOLs), 6 mm in optical diameter were compared. Best corrected visual acuity was > or = 8/10 (Monoyer chart) Parinaud 2 in all cases. Major differences of functional performance in favour of monofocal IOLs were found outside standard conditions of vision (low contrast and illumination levels). A significant difference in contrast sensitivity was found for each spatial frequency in favour of multifocal IOLs (0.0016 < p < 0.05). Mesopic vision was statistically higher in the monofocal IOL group (p = 0.0015). Moreover, dynamic visual acuity allowed accurate evaluation of the difference in performance between these two models of implant. In view of these results multifocal IOLs should be reserved for patients with normal psychosensitive adaptation; an ocular pathology that could alter contrast sensitivity or mesopic vision is a contraindication for multifocal IOLs.
4
Low-Contrast Letter Charts as a Test of Visual Function
D. Regan, David R. Neima · Ophthalmology · 1983 · 210 citations
Food and Drug Administration Study Update
Richard L. Lindstrom · Ophthalmology · 1993 · 99 citations