Publication | Open Access
A dedicated cone‐beam CT system for musculoskeletal extremities imaging: Design, optimization, and initial performance characterization
147
Citations
30
References
2011
Year
The results indicate that the proposed system will deliver volumetric images of the extremities with soft-tissue contrast resolution comparable to diagnostic CT and improved spatial resolution at potentially reduced dose. Cascaded systems analysis provided a useful basis for system design and optimization without costly repeated experimentation. A combined process of design specification, image quality analysis, clinical feedback, and revision yielded a prototype that is now awaiting clinical pilot studies. Potential advantages of the proposed system include reduced space and cost, imaging of load-bearing extremities, and combined volumetric imaging with real-time fluoroscopy and digital radiography.
| Year | Citations | |
|---|---|---|
Page 1