Analytical Letters · 1989 · 21 citations · 17 references
Life SciencesBiochemistryChemistryAnalytical ChemistryElectrochemistryEnzymologyElectroanalytical ChemistryElectrochemical KineticsEnzyme AssaysEnzyme InhibitionEnzyme KineticsRedox EnzymesAmperometric Enzyme BiosensorsChemical Enzyme InhibitorsElectrochemical Enzyme AssaysEnzyme Inhibition AssaysEnzyme Inhibition KineticsGlucose OxidaseMetal Ion InhibitionOxidase Catalysis
Abstract A flow system incorporating an amperometric glucose oxidase enzyme electrode has been used to study the inhibitory effects of 16 metal cations on glucose oxidase. Only copper(II), mercury(II) and silver(I) caused any significant inhibition. the enzyme electrode could be reactivated by EDTA, the reactivation being most effective for copper(II) and least so for silver(I). Other complexing agents were tried for reactivation but proved to be unsatisfactory. The ability to reactivate the enzyme on the electrode following copper(II) inhibition, and the linear response of the system to the level of this inhibitor according to I/A = -9.49 × 10−7 log([Cu]/M) + 4.84 × 10−8; r = 0.994 between 2.5 × 10−4M and 5 × 10−3M [Cu]2+ indicates a prospect for the use of a flow system for determining enzyme inhibitors in samples.
17
Textbook of quantitative inorganic analysis
Journal of the Franklin Institute · 1953 · 2.1K citations
Ferrocene-mediated enzyme electrode for amperometric determination of glucose
Anthony E. G. Cass, G.B. Davis, Graeme D. Francis et al. · Analytical Chemistry · 1984 · 1.6K citations
Stuart J. Updike, G. P. Hicks · Nature · 1967 · 1.3K citations
Analytical Biochemistry · 1982 · 780 citations