Monitoring of radiofrequency thermal ablation in liver tissue through fibre Bragg grating sensors array

Daniele Tosi, Edoardo Gino Macchi, Giovanni Braschi, Mario Gallati, Alfredo Cigada, Sven Poeggel, Gabriel Leen, Elfed Lewis

Electronics Letters · 2014 · 58 citations · 3 references

DOIFull text

Open access

Concepts

Abstract

A fibre Bragg grating (FBG) array‐based sensing probe has been installed on a device for radiofrequency thermal ablation (RFTA). The probe is made of five FBGs with 0.5 cm active area and 1 cm spacing, to provide quasi‐distributed thermal pattern measurements. Multiple experiments have been conducted on porcine liver, reporting the temperature pattern along the ablation axis. Thermal maps allow an estimation of the RFTA efficiency and spatial extension in the liver.

References