Journal of Fluids Engineering · 2004 · 88 citations · 33 references
EngineeringBiomedical EngineeringCardiologyComputational Fluid DynamicsFluid DynamicsFluid MechanicsBiofluid MechanicsHemodynamic SupportHemodynamicsHemodynamics ModelingBlood Flow DynamicsCardiovascular Fluid DynamicsComputational HemodynamicsContinuous FlowLeft Ventricular Assist DevicesNumerical Simulation Of StreaklinesShear HistoryShear Stress
This study explores a quantitative evaluation of blood damage that occurs in a continuous flow left ventricular assist device (LVAD) due to fluid stress. Computational fluid dynamics (CFD) analysis is used to track the shear stress history of 388 particle streaklines. The accumulation of shear and exposure time is integrated along the streaklines to evaluate the levels of blood trauma. This analysis, which includes viscous and turbulent stresses, provides a statistical estimate of possible damage to cells flowing through the pump. Since experimental data for hemolysis levels in our LVAD are not available, in vitro normalized index of hemolysis values for clinically available ventricular assist devices were compared to our damage indices. This approach allowed for an order of magnitude comparison between our estimations and experimentally measured hemolysis levels, which resulted in a reasonable correlation. This work ultimately demonstrates that CFD is a convenient and effective approach to analyze the Lagrangian behavior of blood in a heart assist device.
33
Red Blood Cell Damage by Shear Stress
L.B. Leverett, J. D. Hellums, Clarence P. Alfrey et al. · Biophysical Journal · 1972 · 680 citations · Full text
A Couette viscometer for short time shearing of blood
G. Heuser, Robert Opitz · Biorheology · 1980 · 247 citations
Engineering, Physics, Fluid Dynamics +16
SOME MECHANICAL EFFECTS THAT INFLUENCE HEMOLYSIS
Perry L. Blackshear, Frank D. Dorman, Joseph H. Steinbach · ASAIO Journal · 1965 · 184 citations
Engineering, Bioengineering, Surgery +15