High-resolution digital retroillumination imaging of the posterior lens capsule after cataract surgery

Milind Pande, Paul G. Ursell, David J. Spalton, S Kundaiker

Journal of Cataract & Refractive Surgery · 1997 · 86 citations · 15 references

Concepts

Abstract

PURPOSE: To develop a system for high-resolution imaging of the posterior lens capsule after intraocular lens surgery for objective assessment of posterior capsule opacification (PCO). SETTING: Department of Cataract and Refractive Surgery, St. Thomas' Hospital, London, United Kingdom. METHODS: A system was developed that uses coaxial illumination and imaging based on Zeiss components with a digital camera directly linked to a computer for online image verification and image analysis. RESULTS: The system produced high-resolution digital images with even background illumination of sufficient quality to demonstrate progressive lens epithelial cell changes that are amenable to computer image analysis. CONCLUSION: This system produced excellent images for objective documentation and quantitative measurement of PCO.

References

15