MRI-based treatment planning with pseudo CT generated through atlas registration
182
Citations
25
References
2014
Year
MRI-based treatment planning with pseudo CTs generated through atlas registration is feasible for pediatric brain tumor patients. The doses calculated from pseudo CTs agreed well with those from real CTs, showing dosimetric accuracy within 2% for the PTV when multiple atlases were used. The arithmetic mean process may be a reasonable choice over PRGP for the synthesis scheme considering performance and computational costs.