Analytical Methods · 2013 · 39 citations · 54 references
PharmacyChemistryToxicologyAnalytical ChemistryAnalytical ToxicologyMass Spectrometric AnalysisContaminant AnalysisInstrumental AnalysisIonization TechniquesMass SpectrometryElectrospray IonizationIonizationMass Spectrometric Drug AnalysisMass Spectrometry Characterization
Direct mass spectrometry analysis of untreated samples of volumes as low as 0.2 µL were achieved using fast extraction and nanoESI (electrospray ionization) in a combined fashion. The analytes in dried samples on paper substrates were extracted by organic solvent in a nanoESI tube and ionized with a high voltage applied for generating a spray. The ionization source produced stable signals for different atmospheric pressure interfaces of triple quadrupole instruments. Analysis time more than 20 minutes were available with 10 µL solvent consumed for the entire analysis process. The performance in qualitative and quantitative analysis was characterized with a wide variety of samples. Limits of detection as low as 0.1 ng/mL (corresponding to an absolute amount of 0.05 pg) were obtained for analysis of atrazine in river water, thiabendazole in orange homogenate, and methamphetamine in blood.
54
Analytical Properties of the Nanoelectrospray Ion Source
Matthias Wilm, Matthias Mann · Analytical Chemistry · 1996 · 1.9K citations