Publication | Closed Access
Pressure Relief Characteristics of an Alternating Pressure System
10
Citations
0
References
1990
Year
Numerical SimulationTraditional Hospital MattressFluid PowerHospital MattressPressure VesselMedicineOrthopaedic SurgeryPressure Relief CharacteristicsShock CompressionAnesthesiologyAlternating Pressure Mattress
The pressure relieving characteristics of an alternating pressure mattress are compared to the characteristics of a traditional hospital mattress. The trochanteric pressures of 20 healthy volunteers had a mean of 95 (SD = 49) on a hospital mattress while the trochanteric pressures on the alternating pressure mattress had a mean of 67 (SD = 9) for the inflated cells and a mean of 23 (SD = 7) for the deflated cells. Statistical analysis revealed significantly lower interface pressure for the alternating pressure mattress.