The Atmospheric Research and Exposure Assessment Laboratory (AREAL) of the US Environmental Protection Agency at Research Triangle Park, North Carolina, has studied the Tedlar bag for the applicability of this material in sampling different organic compounds and has designed a Tedlar Bag Sampling System which reduces the sampling rate from 1-1.5 L/min to 50 mL/min to make the sample volume more manageable. This new, smaller version of the original Tedlar bag sampling system was designed to meet the following criteria: It must hold a negative pressure of 10 in. of water to extract emission air samples from a negative-pressure stack; The bag must be visible so that the operator can see it inflate during sampling; It must be small enough to be easily handled; and It must not show any bias compared to the standard sampling system which collects the samples at a much higher rate. This system was designed at AREAL and tested in the field side by side with the conventional large bag sampling system for possible bias of the small bag in sampling.