The LNE-LNHB water calorimeter for primary measurement of absorbed dose at low depth in water: application to medium-energy x-rays

25

Citations

18

References

2013

Year

Abstract

Water calorimeters are used to establish absorbed dose standards in several national metrology laboratories involved in ionizing radiation dosimetry. These calorimeters have been first used in high-energy photons of (60)Co or accelerator beams, where the depth of measurement in water is large (5 or 10 cm). The LNE-LNHB laboratory has developed a specific calorimeter which makes measurements at low depth in water (down to 0.5 cm) easier, in order to fulfil the reference conditions required by the international dosimetry protocols for medium-energy x-rays. This new calorimeter was first used to measure the absorbed dose rate in water at a depth of 2 cm for six medium-energy x-ray reference beams with a tube potential from 80 to 300 kV. The relative combined standard uncertainty obtained on the absorbed dose rate to water is lower than 0.8%. An overview of the design of the calorimeter is given, followed by a detailed description of the calculation of the correction factors and the calorimetric measurements.