Optimization of x-ray spectra in digital mammography through Monte Carlo simulations

Diego Merigue da Cunha, Alessandra Tomal, M.E. Poletti

Physics in Medicine and Biology · 2012 · 28 citations · 47 references

DOIFull text

Open access

Concepts

Abstract

In this work, a Monte Carlo code was used to investigate the performance of different x-ray spectra in digital mammography, through a figure of merit (FOM), defined as FOM = CNR²/(¯)D(g), with CNR being the contrast-to-noise ratio in image and [Formula: see text] being the average glandular dose. The FOM was studied for breasts with different thicknesses t (2 cm ≤ t ≤ 8 cm) and glandular contents (25%, 50% and 75% glandularity). The anode/filter combinations evaluated were those traditionally employed in mammography (Mo/Mo, Mo/Rh, Rh/Rh), and a W anode combined with Al or K-edge filters (Zr, Mo, Rh, Pd, Ag, Cd, Sn), for tube potentials between 22 and 34 kVp. Results show that the W anode combined with K-edge filters provides higher values of FOM for all breast thicknesses investigated. Nevertheless, the most suitable filter and tube potential depend on the breast thickness, and for t ≥ 6 cm, they also depend on breast glandularity. Particularly for thick and dense breasts, a W anode combined with K-edge filters can greatly improve the digital technique, with the values of FOM up to 200% greater than that obtained with the anode/filter combinations and tube potentials traditionally employed in mammography. For breasts with t < 4 cm, a general good performance was obtained with the W anode combined with 60 μm of the Mo filter at 24-25 kVp, while 60 μm of the Pd filter provided a general good performance at 24-26 kVp for t = 4 cm, and at 28-30 and 29-31 kVp for t = 6 and 8 cm, respectively.

References

47