Radiation Protection Dosimetry · 2006 · 20 citations · 10 references
MedicineNuclear EngineeringRadiation OncologyNuclear PhysicsRadiation DosimetryRadiation PhysicsAbsorbed DoseMonte Carlo DosimetryMonte Carlo Radiation TransportRadiation TherapyRadiobiological ModelingTreatment Planning PhysicsDose Distribution CalculationsMonte Carlo Dose SimulationMonte Carlo Neutron TransportMonte Carlo Radiobiological SimulationMonte Carlo Transport SimulationNeutron FluxNeutron Flux SpectrumRoom Design
In this work, the behaviour of scattered and thermal photoneutron fluences in a radiotherapy treatment room was investigated by means of Monte Carlo simulation. The MCNP code was used to study the dependence of these neutron fluences on room design, room area and the effect of neutron moderator materials placed on the room walls. The results of the investigation showed a poor agreement between the simulations and empirical approximations, suggesting that the formulae found in the literature can underestimate the neutron flux inside a radiotherapy room.
10
Analysis of photoneutron spectra produced in medical accelerators
C. Ongaro, Alba Zanini, U. Nastasi et al. · Physics in Medicine and Biology · 2000 · 102 citations
Investigation of secondary neutron dose for dynamic MLC IMRT delivery
Rebecca M. Howell, Michele Ferenci, Nolan E. Hertel et al. · Medical Physics · 2005 · 84 citations · Full text
A study of neutron spectra from medical linear accelerators
A. Facure, R. C. Falcão, A.X. Silva et al. · Applied Radiation and Isotopes · 2004 · 63 citations