Dose corrections for low‐density tissue inhomogeneities and air channels for 10‐MV x rays

Melissa Young, R. O. Kornelsen

Medical Physics · 1983 · 65 citations · 0 references

Concepts

Abstract

Methods of correcting for tissue inhomogeneities which consider only changes in the photon fluence are of limited usefulness for 10-MV x rays. Although there is normally transient electronic equilibrium on the central axis beyond the depth of maximum buildup in soft tissue, when the beam enters a low-density material the beam profile is degraded and there is a loss of lateral electron equilibrium which reduces the dose both within and beyond the inhomogeneity. This paper describes experimental measurements of the dose within and near large, low-density inhomogeneities simulating lungs and also behind air channels (where similar effects occur). Methods of calculating correction factors are discussed.