Anesthesia & Analgesia · 1996 · 37 citations · 9 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryMicrosurgeryOrthopedicsShoulder SurgeryCompartment SyndromeCompartment Syndrome MonitoringElbow FracturesElbow InjuriesShoulder InjuriesUpper Limb OrthopedicsCompartment Syndrome AssessmentCompartment Syndrome ManagementFracture-associated Compartment SyndromeMicrovascular Bone ReconstructionMicrovascular Bone TransferNoninvasive Blood Pressure MeasurementTibia FracturesTibial Fracture Compartment Monitoring
We describe a patient who developed biceps compartment syndrome due to malposition of a noninvasive blood pressure (NIBP) cuff across the elbow with subsequent flexion of the joint. Case Report A 29-yr-old man sustained a tibia-fibula fracture with extensive bone loss after a motorcycle accident. He had no past medical history and his vital signs were normal. He was mesomorphic, 168 cm tall, weighed 70 kg, and his physical examination was unremarkable. The patient underwent a microvascular free fibular bone transfer of the left fibula to the right tibia. Prior to surgery, a 125-mm wide NIBP cuff was wrapped firmly around the upper left arm. Cuff size was determined by visual inspection to be approximately 40% of the midarm circumference. The NIBP was cycled every 5 min (Hewlett Packard Component Monitoring System NIBP module 1008-B; Hewlett-Packard, Palo Alto, CA). A right arm vein was cannulated with a 16-gauge catheter. General anesthesia was induced with intravenous midazolam, droperidol, fentanyl, d-tubocurarine, thiopental, and pancuronium and the trachea was intubated via direct laryngoscopy. Sixty minutes after induction of anesthesia, a 20-gauge right radial arterial catheter was inserted, and the NIBP was cycled every 30 min. He was placed in the left lateral decubitus position where he remained for 9 h. The nondependent left arm was supported by a chest roll in a 60 degrees semiflexed position. The NIBP tubing was unobstructed. Anesthesia was maintained with O2, N2 O, isoflurane, fentanyl, and pancuronium. The blood pressure (BP) varied between 110/60 and 120/70 mm Hg, except for 10 min when it reached a minimum of 80/60 mm Hg. He received 8800 mL of crystalloid, 7 U of packed red blood cells, and 1000 mL of 5% albumin; blood loss was 2000 mL; urine output 1740 mL. The initial hematocrit was 40%, reached a minimum of 24%, and was 30% at the end of surgery. Arterial blood gases and electrolytes were normal. Muscle relaxation was reversed with neostigmine and glycopyrrolate but the trachea remained intubated because of residual anesthesia. In the recovery room, the patient was breathing spontaneously via a T-piece, awake and alert with normal vital signs. He repeatedly flexed and extended his left fingers and with questioning, indicated left upper arm pain exacerbated by active or passive biceps extension. A left digit pulse oximeter indicated 100% saturation. Swelling, erythema, induration, and tenderness was noted over the left biceps area and edema extended to the wrist. The pattern of skin erythema, impressioning, and blistering at the BP cuff site indicated the cuff had been applied across the antecubital fossa (Figure 1). Pinch sensation, grip strength, and finger and wrist extension of the left hand were diminished. Capillary refill was good and radial and ulnar arterial pulses were palpable. Biceps compartment pressures were 46 mm Hg in the left and 12 mm Hg in the right.Figure 1: Induration (I), erythema (E), and blistering (B) of the skin over the distal brachium and proximal antebrachium indicating that the blood pressure cuff was positioned across the antecubital fossa.The patient was brought to a different operating room and anterior and posterior upper arm fasciotomies were performed under general anesthesia. The biceps muscle was edematous, but viable and no hematoma was found. The triceps muscle appeared normal. The NIBP cuff and machine used during the first surgery was tested and functioned properly. The day after fasciotomy, the biceps muscle was more edematous. There was decreased sensation in the distribution of the lateral and posterior cutaneous nerves of the forearm and the dorsal aspect of the thumb. He had full range of motion but significant weakness of grasp, thumb opposition, and finger and wrist extension. Eight days after fasciotomy, muscle edema subsided and the fasciotomy wounds were closed primarily. Sensory and motor neuropathy of the distal median and radial nerves resolved almost completely at the time of discharge on postoperative Day 23. The patient recovered full neurologic function 8 wk after surgery. Discussion In the two previous reports describing compartment syndrome after use of a NIBP device, it was proposed that excessive machine cycling or device malfunction was the cause of this complication [1,2]. In our patient, frequent cuff cycling or excessive cycling pressures are unlikely factors. We propose that increased biceps compartment pressure for nine hours due to firm application of the deflated cuff across the antecubital fossa with subsequent arm flexion to 60 degrees caused compartment syndrome. This mechanism of compartment syndrome is similar to that observed after prolonged application of a circumferential pneumatic compression device [3]. Peripheral polyneuropathy was likely caused by nerve ischemia due to NIBP cuff malposition and/or compartment syndrome. In fact, mild nerve injury is probably not rare after long surgical cases, especially when a NIBP is frequently cycled. Hynynen et al. [4] observed transient hand numbness and weakness in the same arm used to measure NIBP in 2 of 22 patients who underwent a microsurgical procedure lasting a mean of eight hours. The NIBP device is very safe [5] but has on rare occasion been associated with compartment syndrome [1,2], neuropathy [6-8], and less serious problems such as thrombophlebitis, venous stasis, petechiae, ecchymoses, and phlebitis [9-12]. The incidence of these complications is unknown. Suggestions for proper use of the NIBP device include [1,2,4,6,7,11,13,14]: 1) do not wrap the cuff tightly; 2) do not apply the cuff across a joint, bony prominence, or superficial nerve (especially peroneal and ulnar nerves); 3) select the maximum NIBP cycle time consistent with safe monitoring; 4) inspect the cuff site and extremity periodically during prolonged cuff application, especially when the patient is at the extremes of age or has a depressed level of consciousness; for a neonate who requires chronic monitoring, consider removing the NIBP cuff between measurements; 5) record cuff location and cycle time; 6) keep device alarms enabled; and 7) consider the possibility of a device malfunction. Especially when used for prolonged periods, the oscillometric NIBP device should be used with care in patients with depressed consciousness, neuropathy [8], irregular cardiac rhythm (HP Component Monitoring System Users Guide, Hewlett Packard, 1990), labile high BP [2], increased arm activity, (muscular straining, shivering, movement disorders, or seizures), and arterial insufficiency. The oscillometric NIBP cannot be used during cardiac bypass or cardiac arrest and may be less accurate when there is severe hypotension or shock [15]. In these latter situations, and during prolonged surgical procedures, it may be best to use an arterial pressure monitor [2].
9
Hazard of Automatic Noninvasive Blood Pressure Monitoring
A. P. Showman, Eugene K. Betts · Anesthesiology · 1981 · 30 citations