Concepedia

Publication | Closed Access

Evaluation of volumetric modulated arc therapy for cranial radiosurgery using multiple noncoplanar arcs

110

Citations

15

References

2011

Year

Abstract

For the aforementioned planning and delivery system and cranial lesions greater than 7 mm in diameter, multiple noncoplanar arc VMAT consistently provides accurate and high quality cranial radiosurgery dose distributions with low doses to healthy brain tissue and high dose conformity to the target. These qualities may make multiple noncoplanar arc VMAT suitable for a greater range of prescription doses or larger and more irregular lesions. For smaller and/or rounder lesions there are other clinically acceptable treatment techniques that may involve fewer couch angles or arcs and reduce treatment times.

References

YearCitations

Page 1