IEEE Transactions on Instrumentation and Measurement · 1985 · 20 citations · 1 references
EngineeringElectrical EngineeringInstrumentationMetrologyUncertainty QuantificationAnalog InstrumentationCalibrationElectrical MeasurementMeasurement In ScienceMeasurement StandardsMeasurement UncertaintyElectrical Measurement SystemsMeasurement AccuracyPrecision MeasurementReference Voltage AccuracyVoltage MeasurementVoltage Measurement CircuitsVoltage Reference DesignVoltage References
An automated measuring system has been developed for measuring arbitrary-voltage references in the range of 1-10 V with an inherent measurement accuracy of ± 0.22 ppm (3¿). This paper discusses the design and uncertainty analysis of the system and presents data obtained on a Zener voltage reference.
1