Preconcentration of Cd(II), Pb(II), Co(II), Ni(II), and Cu(II) by solid-phase extraction method using 1,10-phenanthroline

Bożena Puzio, Barbara Feist

Journal of Analytical Chemistry · 2009 · 20 citations · 20 references

Concepts

Abstract

Solid-phase extraction (SPE) method for preconcentration and determination of Cd(II), Pb(II), Co(II), Ni(II), and Cu(II) aqueous samples by inductively coupled plasma optical emission spectrometry is described. The preconcentration of analytes is accomplished by retention of their chelates with 1.10-phenanthroline in aqueous solution on a solid phase containing carboxylic acid (COOH) bonded to silica gel in a column. The limits of detection values (defined as “3s” where “s” is standard deviation of the blank determination) are 3.6 μg/L for Cd(II), 17.5 μg/L for Pb(II), 3.1 μg/L for Co(II), 2.1 μg/L for Ni(II), and 4.4 μg/L for Cu(II) and corresponding limit of quantification (6s) values are 7.2, 35, 6.2, 4.2 and 8.8 μg/L, respectively. As a result, a simple method was elaborated for the group concentration and determination of the above mentioned metals in reference material and in samples of plant material.

References

20