EngineeringNuclear EngineeringPhysicsNuclear PhysicsRadiation DetectionRadiation MeasurementDetector TechnologyDosimetry InstrumentationNeutron DetectionNuclear FusionRadiation DetectorsDetector Efficiency CalibrationFast Neutron DetectionNeutron Detector CalibrationNeutron Detector EfficiencyNeutron DetectorsNeutron DiagnosticsRadiators
We have measured the efficiency (tracks per incident neutron) of pure CR-39 for detecting DD and DT neutrons. Neutrons having average energies of 2.9 MeV (DD) and 14.8 MeV (DT) were produced by a 200-keV electrostatic accelerator and the neutron yields were measured using the associated particle counting technique. All CR-39 samples irradiated by DD or DT neutrons were etched for 2 h in a 70°, 6.25-N_ NaOH bath. For bare CR-39, the efficiencies for detecting 2.9- and 14.8-MeV neutrons were found to be (1.3±0.4)×10−4 and (5.0±1.8)×10−5, respectively. We also investigated using CR-39 and polyimide as proton radiators. For detecting 2.9-MeV neutrons, the radiators had no significant effect on efficiency; but for detecting 14.8-MeV neutrons the polyimide radiator increased the efficiency to (7.8±2.8)×10−5.