Comparison of Stiffness and Failure Load of Two Cervical Spine Fixation Techniques in an In Vitro Human Model

13

Citations

15

References

2005

Year

Abstract

This study demonstrates that a plated Smith-Robinson cervical discectomy and fusion construct provides greater stiffness and failure load and reduced ROM across operated levels than a single interbody cage construct. Although clinical success may not directly correlate with biomechanical data, these results raise concern regarding the use of a single threaded interbody cage as a stand-alone device for cervical interbody fusion.