Based on these data, it is clear that both the nail systems showed a higher biomechanical stability with a lower TAD. The 2 specimens that failed with the Intertan in the cyclic tests had a TAD ≥49 mm. The cutout failures that we detected during cyclic testing in the Gamma system had a TAD ≥30 mm. Thus, it is clear that the TAD affects failure independent of the implant used. With a less than ideal lag screw placement, however, the Intertan system with 2 integrated screws was able to withstand higher loads in this study.