Publication | Closed Access
Analysis of Factors Associated with Volumetric Data Errors in Gamma Knife Radiosurgery
28
Citations
5
References
2008
Year
Volume discrepancies existed between those volumes computed by the PGK-WS and volumes determined by laser scanning. The volumetric data errors were reduced through the acquisition of more slices through the phantom and a more spherical morphology of the phantom. Relatively few system volume errors were observed between those by the PGK-WS and PACS except for a significant discrepancy for the irregular surface phantom. For the rectangular-shaped phantom, the volumetric data errors were significantly related to slice orientation of measurement. When measuring the tumor response in GK radiosurgery or follow-up, an error of as large as 20% is possible for irregularly shaped object and with MRIs using <or=5 slices through the region of interest.
| Year | Citations | |
|---|---|---|
Page 1