Urine particle evaluation: a comparison between the UF-1000<i>i</i> and quantitative microscopy

Fabio Manoni, Agostino Tinello, Lucia Fornasiero, Paolo Hoffer, Valeria Temporin, Sara Valverde, Gianluca Gessoni

Clinical Chemistry and Laboratory Medicine (CCLM) · 2010 · 61 citations · 11 references

Concepts

Abstract

BACKGROUND: The study of urine particles plays a key role in the diagnosis of kidney diseases. In this study, the authors evaluated the correlation between the UF-1000i and quantitative manual microscopy. METHODS: A total of 214 untreated urine samples were studied using the Sysmex UF-1000i and compared with results obtained from quantitative manual microscopy using the Fuchs-Rosenthal counting chamber. RESULTS: Using Pearson statistics, we observed satisfactory correlation between the UF-1000i and quantitative microscopy: for red blood cells (RBCs) r was 0.98, for white blood cells (WBCs) r was 1.00, for epithelial cells (EC) r was 0.96, and for casts r was 0.69. Using linear regression statistics, we also observed satisfactory correlation between the UF-1000i and quantitative microscopy: for RBCs R(2) was 0.95, for WBCs R(2) was 0.99, for EC R(2) was 0.92, and for casts R(2) was 0.48. CONCLUSIONS: In our experience, automated urine particle analysis performed using the Sysmex UF-1000i analyzer is sufficiently precise and improves the workflow in a routine laboratory. Precision was satisfactory and concordance with the reference method is good for RBC, WBC and EC; for casts microscopic observation is required for flagged samples to discriminate hyaline from pathologic casts.