Analytical Chemistry · 2003 · 129 citations · 27 references
Natural SciencesChemistryAnalytical ChemistryElectroanalytical MethodsElectrochemistryElectroanalytical ChemistryElectrochemical DetectionElectrochemical SensorsMicrofluidic BiosensorsMicrofluidic DevicesMicrofluidicsCapillary ElectrophoresisCapillary Electrophoresis Method DevelopmentElectrochemiluminescent BiosensorsElectrode DesignElectrophoretic DetectionLinear RangeMicrochip ElectrophoresisPhotolithographic Patterning
This paper describes an indium tin oxide (ITO) electrode-based Ru(bpy)3(2+) electrochemiluminecence (ECL) detector for a microchip capillary electrophoresis (CE). The microchip CE-ECL system described in this article consists of a poly(dimethylsiloxane) (PDMS) layer containing separation and injection channels and an electrode plate with an ITO electrode fabricated by a photolithographic method. The PDMS layer was reversibly bound to the ITO electrode plate, which greatly simplified the alignment of the separation channel with the working electrode and enhanced the photon-capturing efficiency. In our study, the high separation electric field had no significant influence on the ECL detector, and decouplers for isolating the separation electric field were not needed in the microchip CE-ECL system. The ITO electrodes employed in the experiments displayed good durability and stability in the analytical procedures. Proline was selected to perform the microchip device with a limit of detection of 1.2 microM (S/N = 3) and a linear range from 5 to 600 microM.
27
Carlo S. Effenhauser, Gerard Bruin, Aran Paulus et al. · Analytical Chemistry · 1997 · 639 citations
Timothy J. Johnson, David Ross, Laurie E. Locascio · Analytical Chemistry · 2001 · 546 citations