Concepedia

Publication | Closed Access

Evidence for Many Resolvable Structures within Conformation Types of Electrosprayed Ubiquitin Ions

173

Citations

40

References

2006

Year

Abstract

A new two-dimensional ion mobility spectrometry approach combined with mass spectrometry has been used to examine ubiquitin ions in the gas phase. In this approach ions are separated in an initial drift tube into conformation types (defined by their collision cross sections) and then a gate is used to introduce a narrow distribution of mobility-separated ions into a second drift tube for subsequent separation. The results show that upon selection a narrow peak shape is retained through the second drift tube. This requires that at 300 K the selected distribution does not interconvert substantially within the broader range of structures associated with the conformation type within the approximately 10-20 ms time scale of these experiments. For the [M + 7H]7+ ion, it appears that many ( approximately 5-10) narrow selections can be made across each of the compact, partially-folded, and elongated conformer types, defined previously (Int. J. Mass Spectrom. 1999, 187, 37-47).

References

YearCitations

Page 1