Publication | Closed Access
A comparison of hemodynamic parameters derived from transthoracic electrical bioimpedance with those parameters obtained by thermodilution and ventricular angiography
87
Citations
29
References
1997
Year
The 95% confidence range defining the limits of agreement between cardiac output and volumetric data estimated by bioimpedance, with the cardiac output measurement by thermodilution and the volumetric data estimated from left ventriculography, were wide, making the degree of agreement clinically unacceptable. In the opinion of the authors, impedance cardiography should not replace invasive hemodynamic monitoring at this time.