Analytical Letters · 1993 · 17 citations · 20 references
EngineeringBiochemistryFood ScienceBiosensorsElectroanalytical MethodsFood ChemistryColor AnalysisEnzymatic BiosensorsEnzyme-based BiosensorsOptical BiosensorsOptical SensorsSugar AnalysisGlucose ContentGlucose OxidaseGlucose QuantificationGlucose SensingHydrogen Peroxide GenerationOxidase CatalysisPrecisionRelative Precision
Abstract A flow injection system was developed for the determination of glucose based on fiber optic chemiluminescence measurements. The hydrogen peroxide produced from the glucose oxidase catalyzed reaction was quantified by measuring the intensity of luminol chemiluminescence in the presence of excess ferricyanide. Glucose oxidase was immobilized on aminopropyl glass using glutaraldehyde and packed in a reactor. The chemiluminescence was transported to the detector using an optical fiber. The system responds linearly to glucose in the concentration range 0.20 mM up to 2.5 mM, with a relative precision of 5%. Several fruit juices were analyzed for their glucose content and the results were compared with a standard AOAC procedure.
20
Fiber-Optic Sensors for Biomedical Applications
John I. Peterson, Gerald G. Vurek · Science · 1984 · 299 citations
Two-dimensional enzyme electrode sensor for glucose
David Gough, Joseph Y. Lucisano, Pius H. S. Tse · Analytical Chemistry · 1985 · 222 citations
Oxygen optrode for use in a fiber-optic glucose biosensor
María C. Moreno‐Bondi, Otto S. Wolfbeis, Marc J. P. Leiner et al. · Analytical Chemistry · 1990 · 150 citations