Sensorless Controlling Method for a Continuous Flow Left Ventricular Assist Device

45

Citations

14

References

2000

Year

Abstract

We originated a novel control strategy for a continuous flow left ventricular assist device (LVAD). We examined our method by acute animal experiments to change the left ventricular (LV) contractility or LV end-diastolic pressure (LVEDP). To estimate the pump pulsatility without any specific sensor, we calculated the index of current amplitude (ICA) from motor current waveform. The ICA had a peak point (t-i point) that corresponded closely with the turning point from partial to total assistance, and a trough (s-i point) that corresponded with the beginning point of ventricular collapse. The pump flow at the t-i point (Qt-i) had no component of flow regurgitation. In the evaluation of the effects of preload LVEDP, afterload (mAoP), and contractility (max LV dp/dt), we found that preload was the only parameter that significantly influenced Qt-i. We concluded that our method could well control continuous flow LVAD by preventing reversed flow and ventricular collapse.